Instructions
NOTES:
• See the instructions for use supplied with the
battery charger to properly charge the battery
pack before each use.
• For maximum capacity, charge the battery
pack before sterilization, even if the battery
pack has not been used.
• A typical charging sequence will take between
5 and 30 minutes. Charging a fully expended
large battery pack may take up to 75 minutes.
• For maximum capacity, store the battery
pack on the battery charger to maintain a
full charge. DO NOT store the battery pack
installed in a handpiece. Discharge will occur
even though the handpiece is not operating.
• Stryker recommends using battery packs that
have been charged within the last 10 days.
• See the instructions for use supplied with the
handpiece to properly install and remove the
battery pack from the handpiece.
www.stryker.com
• Typical battery pack run time is 20 minutes
or longer, depending on surgical procedure or
technique.
• When the indicator light blinks, reducing the
load on the handpiece may extend the usage
time of the battery pack.

Care Instructions

For processing instructions, see the care
instructions manual supplied with the handpiece.
Disposal/Recycle
See the care instructions manual supplied with
the handpiece.
Per Directive 2006/66/EC, collect
and recycle batteries/battery
packs according to EU Member
State regulations.
This rechargeable battery pack
contains Lithium-Ion.