Do not use in a magnetic resonance environment.
There is a risk of ignition of gases if the device is operated in the presence of
flammable mixtures or mixtures of medicinal products and air or oxygen or
nitrous oxide!
The device must not be operated in rooms in which flammable mixtures or mix-
tures of pharmaceuticals and air or oxygen or nitrous oxide are present, e.g.
operating theatres.
Electric shock!
The housing of the ri-scope L may only be opened by authorised persons.
Damage to the device due to a fall or strong ESD influence!
If the device is not functioning, it must be returned to the manufacturer for
repair.
The device must be used in a controlled environment.
The device must not be exposed to harsh environmental conditions.
Caution!
The caution symbol indicates a potentially dangerous situation that can lead
to minor or moderate injuries. It can also be used to warn of unsafe practices.
The flawless and safe function of ri-scope L instruments is only
guaranteed if original parts and accessories from Riester are used.
Old electronic devices must be disposed of in accordance with the institutional
guidelines for the disposal of expired devices.
Cleaning frequency and sequence must comply with the cleaning regulations of
non-sterile products in the respective facility. Cleaning/disinfection instructions
in the instructions for use must be observed.
We recommend removing the (rechargeable) batteries from the battery handle
before cleaning or disinfecting.
Clean and disinfect the ri-scope L devices carefully so that no liquid penetrates
the interior.
Never place the devices in liquids!
The instruments with battery handles are supplied in a non-sterile condition.
Do not use ethylene, oxide gas, heat, autoclaves or any other harsh methods to
sterilise the device. The devices have not been approved for mechanical repro-
cessing or sterilisation. This leads to irreparable damage!
The patient is not the intended operator.
The product may only be used by qualified personnel.
Qualified personnel are doctors or nurses in hospitals, medical facilities, clinics
and medical practices.