FAULT
Handpiece doesn't work
There is no spray when getting through
the electricity
There is no water when getting through
the electricity
Low vibration
The handpiece generates heat
There is water seeping from the coupling
between the handpiece and the cable
If the problem still can't be solved, please contact with local dealer or manufacturer.
Warnings
The use of ultrasound cleaning device and strong cleaning and disinfection fluids (alkaline pH>9 or acid pH <5) can reduce the
life span of products. The manufacturer takes no responsibility in such cases.
This device shall not be exposed to high temperature above 138 ºC .
Processing limit
The products have been designed for a large number of sterilization cycles. The materials used in manufacture were selected
accordingly. However with every renewed preparation for use, thermal and chemical stresses will result in ageing of the
products. The allowed maximum times of sterilization for handpiece is 600 times.
Initial processing
Processing principles
It is only possible to carry out effective sterilization after the completion of effective cleaning and disinfection. Please ensure
that, as part of your responsibility for the sterility of products during use, only sufficiently validated equipment and product-
specific procedures are uses for cleaning/disinfection and sterilization, and that the validated parameters are adhered to
during every cycle.
Please also observe the applicable legal requirements in your country as well as the hygiene regulations of the hospital or

5. TROUBLESHOOTING

POSSIBLE CAUSES
The tip is loosen
Something wrong with the Handpiece
Something wrong with the Cable
The connect plug between the cable and
circuit board is loosen
The water control switch is vibrates not on
No pressure or low pressure
The water control switch is vibrates not on
There is impurity in the electric-magnetic
valve
Contact our dealers or us
The tip hasn't been screwed on the
handpiece tightly
The tip is shaken loose by vibration
The coupling between the handpiece and
the cable isn't dry
The tip is worn out or deformation
The water control switch is in a low setting
The waterproof "O ring is damaged

6. CLEANING, DISINFECTION AND STERILIZATION

SOLUTIONS
Tighten it
Take out the handpiece and send it back
for repair
Contact our dealers or us
Contact our dealers or us
Turn on the water control switch
Check the presure
Turn on the water control switch
Contact our dealers or us
Clean the water line by three-way syringe
Screw the tips tightly by torque wrench
Screw on the tip tightly
Dry it, especially the water between
handpIece and connector
Change another tip
Turn the water control switch to a higher
grade
Change a new "O " ring
17