Notes: The detachable handpiece is a high-tech device which should be used carefully. Don't use too much torque force to
screw the tip, Screw the tips by torque wrench. In order to the vibration by touch the foot switch by accident, unscrew the
scaling tips after each operation.

Scaling function

Connect the handpiece and the connector fo cable of the main unit of the scaler correctly, and adjust the water and vibration
intensity to a suitable level.
• The handpiece can be handled in the same gesture as a pen in hand.
• Select a suitable scaling tip as you need, screw it on the handpiece tightly by the torque wrench.
• During the clinical treatment, be sure not to make the end of tip touch the teeth vertically.
• Not to make the tip overexert on the surface of the teeth in case of hurting the teeth and damaging the tip.
• After finishing operation, keep the machine working for 30 seconds on the water supply condition in order to clean the
handpiece and the scaling tip.
Endo funtion
Usage process
a) Fix endochuck to handpiece by endo chuck.
b) Unscrew the screw cap on the endochuck.
c) Put the ultrasonic file into the hole in the front of endochuck.
d) Screw down the screw cap with endo wrench to tight up the ultrasonic file.
e) Press option key, turn to endo root canal cleaning function.
f) Press option key, turn the water and vibration intensity to a suitable level.
g) Put the ultrasonic file into the patient's root canal slowly, step on the foot pedal, then make endo root canal cleaning.
During the treatment, turn up the power gradually according to the needs.

Notes

a) When fixing endochuck, it must be screwed down.
b) The screw on the endochuck must be screwed down.
c) Don't press it too much when the ultrasonic file in root canal.
d) Don't start up the foot pedal until the ultrasonic file is in root canal.
e) The power range is supposed from 1st to 5th grades.
f) Please sterilize the endo chuch and handpiece after operation.
Instruction for main components of detachable handpiece
• Nipple: The nipple can be removed, for easy cleaning.
• Handpiece: The main part of the whole handpiece, can be autoclaved underthe high temperature and pressure.
• O ring: Please lubricate the waterproof O ring with dental lube frequently, as sterilication and repeated pulling and inser-
ting will reduce their using life. Change a new one once it is damaged or worn excessively.
• Don't knock or rub the handpiece.
• Don't screw or unscrew the scaling tip and endochuck when step on the foot pedal.
• The handpiece, scaling tip, torque wrench, endo wrench and endochuck must be sterilized before each treatment.
• The internal screw threat of the scaling tips produced by some manufactures is coarse, rusty and collapsed. This will
damage the external screw threat of the handpiece irretrievably. Please use "DTE" brand scaling tips.
• We are only responsible for the safety on the following conditions: The maintenance, repair and modification are made by
the manufacturer or the authorized dealer.