Immunity
Level
test
indicated
by EN
60601-1-2
Conducted RF
3 Veff
EN 61000-4-6
From 150
KHz to 80
MHz
Irradiated RF
3 V/m
EN 61000-4-3
From 80 to
2,5 GHz
emc_cardioline
Conformity
Electromagnetic
level
environment
Portable and mobile RF
communication units must
not be used nearer to any
part of the device,
including its cables, than
the recommended
separation distance
calculated by the equation
applicable to the
transmitter frequency
Recommended separation
distance
d = [3,5/V1]·√P
V1= 3 V
rms
d = [3,5/E1]·vP
E1= 3 V/m
d = [12/E1]·vP
Where P is the max.
nominal output of the
transmitter in Watts (W)
according to the
transmitter manufacturer
and d is the recommended
distance in metres (m).
The field intensity of fixed
RF transmitters, as
determined by an
electromagnetic
investigation of the site,
should be lower than the
conformity level in each
frequency interval.
Interference can be
detected near units
marked with the following
symbol:
Rel. 1.xx – Ed. 1.0
Immunity
®
CARDIOLINE
is an et medical devices SpA brand
w w w . c a r d i o l i n e . b i z
6