The materials used in the manufacture of this device are not suitable for repro-
cessing or resterilisation.
This device is not designed to be reprocessed or resterilised.
Unauthorised re-use or reprocessing
– can cause the device to lose the essential performance properties
intended by the manufacturer.
– leads to a significant risk of cross-infection/ contamination as a result of
potentially inadequate processing methods.
– may cause the device to lose functional properties.
– may cause materials to break down and lead to endotoxic reactions
caused by the residues.
for puncture:
1. Take care to use devices of suitable dimensions (diameter, length), espe-
cially when treating obese patients and children.
2. Use an introducer to insert the SPROTTE
ture incision beforehand on the area where the puncture is to be performed
(blood lancet etc.)
3. Only perform the puncture (even when removing the cannula) with the
stylet in place.
4. To avoid bending or breaking of the cannula, never apply excessive force to
the cannula.
5. If you unexpectedly come into contact with bone, change the direction of
the cannula. Do not try to overcome bone resistance. Failure to adhere to
these rules could cause the cannula to bend or break.
6. Repeated bone contact will damage the cannula tip. On no account you
should continue to use a cannula damaged in this manner. In case of a
previous bone contact remove the cannula (with introduced stylet) and
Introducer in one step.
for injection:
1. Always ensure that the injection site is aseptic.
2. Do not administer drugs that are not indicated for the intended use.
3. Aspirate before the injection of medication. If you observe blood in the cylinder
of the syringe, then the cannula has been introduced improperly. TERMINATE
THE PROCEDURE.
for use with other compatible products:
When using multiple components, familiarise yourself with their operation
before use by checking connections and passages (cannulas, adapters).
cannula and/ or perform a punc-
®