Spezifikationen; How To Make Measurements - PeakTech 1610 Bedienungsanleitung

Digital - zangenmessgeräte
Vorschau ausblenden Andere Handbücher für 1610:
Inhaltsverzeichnis

Werbung

max. Betriebshöhe 3.000 m ü.M.
Abmessungen
80 (B) x 229 (H) x 40 (T) mm
Gewicht
300 g
3.1. Maximal zulässige Eingangswerte
Funktion
A AC, D CA (P 1615, P 1625)
V DC, V AC
Widerstand, Diode, Durchgangs-
prüfung, Frequenz, Arbeitszyklus,
Kapazität
Temperatur (°C/°F)

4. Spezifikationen

4.1. Gleichspannung
Modell
Bereich
Auflösung
P 1610
400mV
100µV
4V
1mV
40V
10mV
400V
100mV
600V
1V
P 1615
400mV
100µV
4V
1mV
40V
10mV
400V
100mV
600V
1V
P 1625
400mV
100µV
4V
1mV
40V
10mV
400V
100mV
600V
1V
Überlastschutz: 600V AC/DC
Max. Eingang
1000 A
600 V DC/AC
250 V DC/AC
60 V DC/24 V AC
Genauigkeit
± 0,8% v.M. + 3 dgt
± 1,5% v.M. + 3 dgt
± 2,0% v.M. + 3 dgt
± 0,8% v.M. + 3 dgt
±1,5% v.M. + 3 dgt
± 2,0% v.M. + 3 dgt
± 0,8% v.M. + 2 dgt
± 1,5% v.M. + 2 dgt
± 2,0% v.M. + 2 dgt
-5-

6. How to make measurements

Before making any measurements read safety precautions. Always
examine the instrument and accessories used with the instrument for
damage, contamination (excessive dirt, grease, etc) and defects.
Examine the test leads for cracked or frayed insulation and make sure
the lead plugs fit snugly into the instrument terminals. If any abnormal
conditions exist, do not attempt to make any measurements.
6.1. Voltage measurements
1. Turn off power to the device under test and discharge all
capacitors.
2. Plug the black test lead into the COM input jack on the meter and
connect the test lead tip to a grounded point (the reference point
for measurement of voltage).
3. Select the desired AC voltage range or DC voltage range. If the
magnitude of the voltage to be measured is unknown always start
with the highest range.
WARNING!
To avoid possible electric shock, instrument damage and/or equipment
damage, do not attempt to take any voltage measurements if the
voltage is above 600 V AC/DC are the maximum voltages that this
instrument is designed to measure. The "COM" terminal potential
should not exceed 600 V measured to ground.
4. Plug the red test lead into the V/Ω-Input jack on the meter and
connect the circuit where a voltage measurement is required.
Voltage is always measured in parallel across a test point.
5. Turn on power the circuit/device to be measured and make the
voltage measurement reduce the range setting if set too high until
a satisfactory reading is obtained.
6. After completing the measurement, turn off power to the
circuit/device under test, discharge all capacitors and disconnect
the meter test leads.