INSTALLATION

8.

USE

 The appliance must be installed and operated in a dry, well ventilated and dust-free area where ambient
temperature is within the range of +5°C to +40°C and relative air humidity does not exceed 70%. Otherwise,
failure-free operation of the compressor cannot be guaranteed. The compressor must be installed so that it
is accessible at all times for operation and maintenance.
accessible.
 The appliance must stand on a flat, sufficiently stable base. See paragraph 5 (Technical data) when
positioning or lifting the compressor.
 Compressors cannot be exposed to outdoor environments. The appliance cannot be used in moist or wet
environments. Do not use the compressor in the presence of explosive gases, dust or combustible liquids.
 Before connecting the compressor to medical equipment, the supplier must confirm that it meets all
requirements for its use. Refer to the technical data of the product for this purpose. When a unit is to be
built-in, classification and evaluation of compatibility must be done by the manufacturer or supplier of the
product to be used.
 Any use other than that described in this manual is not covered by the guarantee, and the manufacturer is
not liable for any damages that may result. The operator/user assumes all risk.
9.
INSTALLATION
Only qualified personnel can install and start up the appliance and train operating
personnel in its correct use and maintenance. Installation and training of all operators
shall be confirmed by the installer's signature on the certificate of installation.
Prior to installation, ensure that the compressor is free of all transport packaging and
stabilizers to avoid any risk of damage to the product.
Caution! When in operation, the compressor is hot. Burns or fire may result if contact is
made by the operator or any flammable material.
Electric cord for connection to electric mains and air hoses may not be broken. The power
cord may not be exposed to pulling, pressure and excessive heat.
9.1.

Placement of the compressor

Fig.6
Handling
Unpacking
09/2014
Please ensure that the appliance label is
- 11 -
NP-DK50 4VR50-3_09-2014

Werbung