Error Code
E 13
E 14
E 15

9.2 NIBP Analysis Errors

NIBP System
Error Code
E 20
E 21
E 22
E 23
E 24
E 25
E 26
E 27
Description
Device reset.
Data corrupted after
reset.
Battery too low to
perform measurements.
Description
Indeterminate systolic
blood pressure (likely
due to inadequate
inflation threshold).
Indeterminate diastolic
pressure (premature
deflation; excessive
motion artifacts or
diastolic below 40
mmHg).
Indeterminate pulse
rate
Systolic blood pressure
under-range (<60
mmHg).
Diastolic blood pressure
under-range (<40
mmHg).
Systolic blood pressure
over-range (>250
mmHg).
Diastolic blood pressure
over-range (>160
mmHg).
Arithmetic overflow in
data calculation.
Corrective Action
Readings in memory, if any, are available
for review. Check device environment for
electro-magnetic interference (ESD,
power interruptions). Connect USB cable
between measurements or when device
is powered off.
Switch off, wait 5 seconds and switch on.
Charge the battery. Operate device on
AC power.
Corrective Action
Check patient for pulse. Check cuff
and re-try.
Check patient for pulse. Check cuff
position and re-try.
Check patient for pulse. Check cuff
position and re-try.
Check patient manually for blood
pressure and pulse.
As for E23 above
As for E23 above
As for E23 above
Repeat measurement.