HEALTHCARE
9.5

Troubleshooting Guide

Troubleshooting
Symptoms
Power is on (green
indicator light is
on), but NIBP
START button
does not initiate
inflation.
Power is on (green
indicator light is
on), but display are
all blank.
Power is on (green
indicator light is
on), and device
LED malfunction '?'
is displayed.
Power is on and
E15 is displayed
when NIBP or
Temp START
button is pressed.
No power (green
indicator light is
Cuff does not
inflate during
operation.
30
Possible Cause(s)
Device is in Automatic Mode
and 6 readings have already
Cuff pressure is still present
Device has deactivated
display after more than 10
minutes of non-operation.
Unrecoverable system error
Power supply not plugged
into wall outlet or unit.
off).
Wall outlet not powered.
Power supply plugged in
properly and power indicator
light does not turn on with
Battery fully discharged.
Undetermined causes.
Cuff or tubing not attached
Leakage in cuff or tubing.
been taken.
occurred.
Battery is low
power switch.
properly.
Corrective Action
Press CLEAR to clear the
memory, and then press
NIBP START to initiate new
BP measurements.
Check pressure in system if
the problem persists.
Return for servicing.
Press any button to restore
display.
Turn power off and ensure
cuff is deflated. Turn power
on again.
Plug Smartsigns® LITE
Series into AC mains supply
and charge the battery.
Ensure unit is correctly set
up.
Have qualified technician
inspect wall outlet and
power breaker.
Re-charge the battery.
Return for servicing.
Check cuff and tubing
Check cuff and tubing for
cracks and tears.
Return for servicing