9.

Error Messages and Trouble Shooting

9.1

NIBP System Errors

1.
If a NIBP system error occurs during operation, an error code will be
displayed in the SYSTOLIC display only.
2.
If an NIBP system error (indicated by '*' in the table below) occurs during
automatic mode, the error code is stored in memory and the series
automatically continues until six measurements are performed.
Occurrence of other NIBP system errors require the user to push NIBP
START to continue the series of automatic measurements or to reset the
device and begin a new series of measurements.
3.
Pressing the review button will show error codes in the SYSTOLIC
display for each reading where an error was detected and stored. The
average will exclude readings where an error was detected.
NIBP System
Error Code
- - -
E 0 *
E 1 *
E 2
E 3
E 4
E 5
E 6
E 7
E 8 *
E 9
E 10
E 11
E 12
26
Description
User-cancelled
measurement.
Too few pulses
detected.
Excessive motion
artifacts (motion
induced errors).
Cuff over-pressure.
Inflation too slow
Inflation too fast
Deflation too slow
Deflation too fast
Internal software fault
Low pulse amplitude
Vacuum detected
Cuff over 300 mmHg for
more than 5 seconds.
Cuff pressurised for
more than 130 seconds.
Automatic cycling
timing violation.
Corrective Action
None required.
Check patient for pulse. Check cuff
and tubing and re-try
Check cuff position and re-try.
Check cuff and tubing and re-try.
Check cuff, tubing and connectors for
leaks/kinks.
As for E3 above
As for E3 above
As for E3 above
Switch off, wait 5 seconds and switch on.
Check patient for pulse. Check cuff
and tubing and re-try
Push clear button. Disconnect cuff and
wait 15 seconds before reconnecting.
Check cuff and tubing and re-try.
Check cuff and tubing for leaks/kinks.
Check connectors for tightness.
Check cuff and tubing for kinks.