Informational, Warning and Error Messages

Information for users displayed in the status display are depicted only temporar-
ily and disappear again after a few seconds.
Depending on the type of message, windows with green, yellow, and red frames
are depicted.
Informational message
Green frame
Warning message
Orange frame
Error message
Red frame
Critical message
Red frame
Informational message
AirSeal is shutting down. System Auto
Calibration in process.
Connect blue tube. Smoke evacuation is
initiating...
Warning message
Check for excessive leakage!
Fluid in filter! Check cannula position! If
fluid level increases Smoke Evacuation
will stop!
Fluid in filter! Check cannula position! If
fluid level increases AirSeal Insufflation
will shut down!
Smoke Evacuation stopped. Insufflation
active.
Smoke Evacuation stopped. Suction line
blocked!
Safety threshold: Pressure >15 mmHg!
Starting... Do not install tube set until
self test is completed!
Occlusion!
Error message
Venting valve error! Restart the device.
If the error occurs again call service!
Electronic defective Restart the device. If
the error occurs again, call service!
Filter contaminated! Insert new tube
set!
Filter contaminated! AirSeal Insuffla-
tion stopped! Insert obturator into
AirSeal cannula to maintain pressure!
Change tube set to continue with
AirSeal function!
Cause
Pressure is slowly reduced. Pump keeps running for a few seconds. No pressure build
up.
Do not turn device off. Wait for calibration.
To enable the smoke gas evacuation function, connect the blue tube to a trocar and
open the valve.
Cause
Long suction or leaks.
Low fluid level of the fluid trap in the
filter housing during Smoke Evacua-
tion mode.
Low fluid level of the fluid trap in the
filter housing during AirSeal mode.
Leaks or low abdominal pressure.
Evacuation blocked.
The set value of 15 is reached and the
user tries to increase the pressure fur-
ther.
Initial self test is "running".
Tube or Veress needle has a blockage:
Cause
Valve malfunction.
Malfunctions in the electronic system.
High fluid level of the fluid trap in the
filter housing
High fluid level of the fluid trap in the
filter housing during AirSeal mode.
Informational, Warning and Error Messages
Status line display
Informational, warning and error messages
Remedy
Check for possible leakage! Use suction
carefully.
Please insert a new tube set.
Please insert a new tube set.
Check for possible leakage! Use suction
carefully
Check the evacuation/suction system for
proper function.
Release the key for 2 seconds. Now you
can set a value up to 20 mmHg.
Wait until the initial self test has com-
pleted.
Localize the cause and open/eliminate the
occlusion.
Remedy
Restart the device. If the error occurs
again, call service!
Restart the device. If the error occurs
again, call service!
Please insert a new tube set.
Please insert a new tube set.