Initial Device Startup

Always check all parts and accessories of the device immediately after receiving
the shipment. The manufacturer considers only replacement claims that have
been immediately submitted or reported to a sales representative or an autho-
rized agent.
Place the device on a flat surface and install in dry environment. The ambient
temperature and humidity must meet the requirements mentioned in chapter
13 "Technical Data", page 43.
Ensure good ventilation of the device. The vents are located on the underside and
on the back (see 5.2 "Device Rear" and 5.3 "Bottom of the Device").
WARNING!
Not explosion-proof
The device is not explosion-proof. Do not use in an area where flammable anes-
thetic gases are present.
CAUTION!
Check to make sure the available mains voltage matches the data listed on the
type label attached to the back of the device. Incorrect voltage can cause errors
and malfunctions and may destroy the device.
Make sure the connection data and technical specifications of the power supply
comply with DIN VDE or national requirements. The mains power supply cable
must be plugged into a properly installed safety wall plug (see DIN VDE 0100-
710). Read the device label located in rear of device (type plate) to determine the
operating voltage of the device.
The power connection must be equipped with a grounding contact. Use the orig-
inal power cable (if included in scope of delivery) to establish a connection be-
tween the mains wall socket and the non-heating device plug located in the rear
of the device.
Only use a certified (UL-listed), removable mains connection cable, type SJT, min-
imal 18 AWG, 3 leads. The plug connectors must comply with NEMA 5-15 or IEC
60320. Grounding will only be reliable if the equipment is connected to a corre-
sponding hospital grade socket.
Integrate the device into the potential equalization system as specified by local
safety rules and regulations.
4.1

Gas Connection

WARNING!
Medically pure CO
2
Make sure to use only medically pure CO
mixtures of gases, high pressure compressed gases, gases with entrapped liq-
uids, or polluted gases must not be used with this device.
Use a high-pressure tube to connect a CO
nection or connect to centralized CO
. Other gases (i.e., helium, N
2
gas cylinder to the rear gas inlet con-
2
gas supply.
2
Delivery inspection
Setting up the device
Mains connection
Grounding contact
Only for U.S. operators
Potential equalization
O, argon),