Herunterladen Inhalt Inhalt Diese Seite drucken

Recommended Safety Distances Between Portable And Mobile Rf Telecommunications Devices And The Airseal® I.f.s - Surgiquest AirSeal i.F.S. Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 57
Electromagnetic Compatibility
EN
40
b
The field strength should be less than 3 V/m for the frequency range of 150 kHz
to 80 MHz.
11.6
Recommended Safety Distances Between Portable and Mobile
RF Telecommunications Devices and the AirSeal® i.F.S.
Recommended safety distances between portable and mobile RF telecommuni-
cations devices and the AirSeal® i.F.S.
The AirSeal® i.F.S. is intended for use in an electromagnetic environment where
HF interferences are controlled. The user/operator of the AirSeal® i.F.S. can con-
tribute to lowering electromagnetic emissions by complying with the mini-
mum distance between portable and mobile HF telecommunications devices
(transmitters) and the AirSeal® i.F.S. - depending on the output power of the
communication device listed below.
Rated output of
Safety distance based on the transmitting frequency [m]
the transmitter
150 kHz to 80 MHz
[W]
d = 1.2P
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
The safety distance d in meters [m] for transmitters with a max. rated output not
listed in the table above can be calculated by applying the corresponding formula
in the respective column. P is the max. rated output of the transmitter in watts
[W] according to the information provided by the manufacturer of the transmit-
ter.
Note 1: The higher frequency range applies to 80 and 800 MHz.
Note 2: These guidelines are probably not realizable in all cases. The distribution
and spread of electromagnetic quantities differs depending on the absorption
and reflection of buildings, objects, and people.
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.2P
d = 2.3P
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23

Werbung