The following gas bottle pressures are displayed:
> 40 bar/580 psi
30 - 40 bar/435 - 580 psi
15 - 30 bar/218-435 psi; Smoke evacuation level will switch to LOW
until tank is replaced.
5 - 15 bar/73-218 psi; warning screen "Gas level low. Prepare to
change gas bottle" appears and audible signals can be heard.
If insufflation is stopped, the warning message "Change gas bottle!"
is displayed and insufflation cannot be restarted until the pressure
is > 15 bar/218 psi.
< 5 bar/73 psi; If insufflation is started, the warning message
"Change gas bottle!" is displayed and acoustic signals (beeps) are
emitted. The gas bottle should be changed immediately.
If insufflation is stopped, the warning message "Change gas bottle!"
is displayed and insufflation cannot be restarted.
While in AirSeal mode, a countdown of 100 s is displayed during
which the empty gas bottle can be changed while maintaining
abdominal pressure.
0 bar/0 psi; If insufflation is started, the warning message "Change
gas bottle!" Insufflation stopped!" is displayed and acoustic signals
(beeps) are emitted. The gas bottle should be changed immediately.
If insufflation is stopped, the warning message "Change gas bottle!"
is displayed and insufflation cannot be restarted.
The following house gas supply pressures are displayed:
House gas supply pressure OK
House gas supply pressure too low
Initial Device Startup
Gas supply with gas bottle
House gas supply
EN
13