Herunterladen Inhalt Inhalt Diese Seite drucken

Window-Meanings - Agilent Technologies TV 301 Navigator Bedienungshandbuch

Inhaltsverzeichnis

Werbung

Window-Meanings

Tab. 8
N.
000
008
100
101
102
103
104
105
106
107
TV 301 Navigator User Manual / 87-900-946-01
RS 232/RS 485 communication description
Window-Meanings
Read/Write
Datatype
R/W
L
R/W
L
R/W
L
R/W
N
R/W
N
R/W
N
R/W
L
R/W
N
R/W
L
R/W
L
Technical Information
Description
Start/Stop (in remote mode
the window is a read only)
Remote (default) or Serial
configuration
Soft Start
(write only in Stop condition)
Set Point type
Set Point threshold (expressed
in Hz, mA or s)
Set Point delay: time between
the pump start and the set
point check (seconds)
Set Point signal activation
type: the signal can be "high
level active" or "low level
active"
Set point hysteresis (in % of
threshold)
Water cooling
Active Stop (write only in
stop)
(M)
17
Admitted Values
Start = 1
Stop = 0
Remote = 1
Serial = 0
(default = 1)
YES = 1
NO = 0
0 = Frequency
1 = Current
2 = Time
(default = 0)
(default = 867)
0 to 99999
(default = 0)
0 = high level
active
1 = low level
active
(default = 0)
0 to 100
(default = 2)
0 = NO
1 = YES
0 = NO
1 = YES
247/282

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis