Classification
Type of protection against electric shock:
Class II
Protection from electric shock is achieved by DOUBLE INSULATION.
Protective earthing or reliance upon installation conditions are not
required.
Degree of protection against electric shock:
Type B
Equipment providing a particular degree of protection against electric
shock, particularly regarding:
1) Allowable leakage current;
2) Reliability of protective earth connection (if present).
Not intended for direct cardiac application.
Method of cleaning and infection control allowed:
Please refer to your Equipment Provider and the VisionAire Service
Manual.
Degree of safety of application in the presence of flammable gases:
Equipment not suitable for such application.
Mode of operation: Continuous duty.