"Specifications continued"
Medical equipment needs special precautions regarding EMC and need to
be installed and put into service according to the EMC information provided in
this section.
Guidance and manufacturer's declaration ± electromagnetic immunity
The unit is intended for use in the electromagnetic environment specified below. The customer or the user of the unit
should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power.
IEC 61000-4-11
Power frequency
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
EN-22
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV line to line
± 2 kV line to earth
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5s
3 A/m
VisionAire™ 2 and VisionAire™ 3
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
Not Applicable
± 1 kV line to line
± 2 kV line to earth
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 s
3 A/m
Electromagnetic environment ±
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the unit
requires continued operation during power
mains interruptions, it is recommended
that the unit be powered from an
uninterruptible power supply (UPS) or
a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
®
AirSep
Corporation
MN157-1 rev. B