9. Replacement of Fuse

Fuse is located in
15
clamp located on the top and bottom of the
15
Fuse Box and pull it out to change the fuse.
(Fig. 8)
When the fuse come down, Insert new fuses into
15
the
Fuse box and then push the
into the original position until hearing a sound
click.
- Fuse: Order Code (120V) : 1202225010 (T2.5AH 250V)
- Fuse: Order Code (230V) : 1202216010 (T1.6AH 250V)
CAUTION
Fuse is burned out when a short circuit occurs or when over-voltage is flowed into the
primary current source. If the cause is uncertain, return the product to an authorized
NSK's service shop for inspection.

10. Maintenance Mode

The unit is provided with a maintenance mode to check the functionality of the switches,
display, foot pedal, motor, etc. While pressing
Cruise Switch at the same time, turn on the Power Switch and keep pressing the button until
beeps are made (for about 2 seconds). At this time, turning the volume from the minimum
position in order displays "oP", "dP", "HL", "Pd" and "in", allowing the following checks. To
release Maintenance Mode, turn Power Switch off and switch on again.
(1) "oP": Switch check
Press the switches on the panel, and the right and/or left lamps will light to check to see if
the switches operate normally.
(2) "dP": Display check
Press
9
Forward/Reverse Selector Switch, and the lamps will light one by one to check to
see if they operate normally. To cancel this check, press
again.
Fuse Box. Release ratchet
Fuse box
15
(operation check)
8
Turbine/Motor Selector Switch and
9
Forward/Reverse Selector Switch
10
Fig.8
Auto
10