Night Cleaner
Night Cleaner
Night Cleaner
Night Cleaner
ENGLISH
LED 4 BLINKING
LED 4
LED 4
LED 4
BLINKING orange
BLINKING
BLINKING
orange
orange
orange
Risk of fire and electric shock.
First disconnect the mains plug.
1° Your device is too hot. Unplug it, wait for 1 hour and start the device again.
2° If error is still present, contact EMS aftersales service.
Note: This error also shows up when the device is operating below the minimum temperature.
In this case, just wait for the device to warm up to ambient temperature.
Water filter leakage
Water filter leakage
Water filter leakage
Water filter leakage
First disconnect the mains plug.
1° Replace the water filter (blue cartridge).
2° If still not solved, contact EMS aftersales service.
Bottle or bottle connection
Bottle or bottle connection leakage
Bottle or bottle connection
Bottle or bottle connection
1° Ensure the bottle cap has been correctly closed.
2° Clean the connection: cap and device sides.
3° Replace the bottle.
4° If still not solved, contact EMS aftersales service.
AIRFLOW
AIRFLOW
AIRFLOW
AIRFLOW
® ® ® ®
connection
connection
connection
connection leakage
leakage
leakage
leakage
1° Ensure the handpiece has been correctly connected to the cord.
2° Clean the interior of the handpiece and the cord terminating end.
®
3° Replace the AIRFLOW
cord gasket as described in paragraph "AIRFLOW® handpiece leakage".
4° If still not solved, contact EMS aftersales service.
I I I I nsufficie
nsufficient or no water
nt or no water from handpiece
from handpiece
nsufficie
nsufficie
nt or no water
nt or no water
from handpiece
from handpiece
1° Make sure you have set your water regulators to 10 (maximum flow on the cord) and verify that the
handpiece is not clogged by removing it and checking the water flow without handpiece.
2° Check your water filter cleanliness and replace it if necessary.
Disconnect the mains plug before servicing any filter.
2° Make sure you have well-connected and sufficient pressure from your water supply.
3° If still not solved, contact EMS aftersales service.
Still some blue liq
Still some blue liquid remaining after rinsing
uid remaining after rinsing
Still some blue liq
Still some blue liq
uid remaining after rinsing
uid remaining after rinsing
1° Make sure you have set your water regulators to 10: maximum flow on the cords.
2° Make sure you have well-connected and sufficient pressure from your water supply.
3° Perform a second rinsing phase before treatment.
4° If still not solved, contact EMS aftersales service.
The unit does not start
The unit does not start
The unit does not start
The unit does not start
1° Check the electrical connection and power socket.
2° Check the fuses at the back of the unit:
First disconnect the mains plug.
Fuses are housed within the power cord socket.
1° Remove the power cord from the device.
2° With the help of a small flat screwdriver, open the fuse-holder cover.
3° Replace fuses only with the exact type required (refer to the "Technical Description" section).
4° If still not solved, contact EMS aftersales service.
Wireless pedal does not work
Wireless pedal does not work
Wireless pedal does not work
Wireless pedal does not work
In the case is evident that the pedal remained pressed for longer than 10min, simply release the pedal
and power cycle the device. If not this case, the problem may have multiple causes. A step-by-step
multiple checks are required:
1° Switch-off the device and disconnect and reconnect both the PIEZON
systems. Try again.
2° Perform a new pairing. This procedure is explained in the paragraph "Pairing a new pedal". Try again.
3° Change the 2x AA lithium batteries and try again.
4° If still not solved, contact EMS aftersales service.