Cleaning
Any part can be cleaned manually or automatically by washer or disinfector. EMS recommends the
use of automatic washer-disinfector (WD) for an optimal effectiveness and part service life.
DO NOT use any Ultra Sound Bath cleaning procedure with the PIEZON
PERIOFLOW
®
handpieces: it may destroy the products.
Manual Cleaning
Manual Cleaning
Manual Cleaning
Manual Cleaning
The following validated process can be used:
Remove away any surface soiling (gross
contamination) on the product with a
wetted cloth and drinking water.
For part having a lumen, flush the lumen
3 times for 5 seconds each using a
water jet pistol.
Immerse in cold tap water for 5 min.
Brush all accessible parts with a soft
bristled nylon brush.
Immerse in 0.5% cleaning solution of
Neodisher
MediClean
Hamburg) at 40°C for 5 min. All surfaces
must be moistened.
Rinse for 60 seconds with a water jet
pistol and rinse thoroughly afterwards
under running tap water.
Fully dry with an air pistol for at least 20
seconds.
Disinfection
Any part can be disinfected manually or automatically by a disinfector. EMS recommends the use
of automatic washer-disinfector (WD) for an optimal effectiveness and part service life.
DO NOT use any Ultra Sound Bath disinfecting procedure with the PIEZON
PERIOFLOW
®
handpieces: it may destroy the products.
Manual Disinfection
Manual
Manual
Manual
Disinfection
Disinfection
Disinfection
ASP CIDEX OPA® 0.55% disinfectant
solution shall be used not diluted and
within
its
Use-Life
ENGLISH
(Dr.
Weigert,
and
Shelf-Life,
FB-618/EN – Rev. A-01 - 2017.03.30
Automated Cleaning
Automated Cleaning
Automated Cleaning
Automated Cleaning
Washer-disinfector must have suitable baskets
to hold small fragile products and has rinsing
connections for the attachment to product
lumen.
Manual pre-cleaning is required and the
following validated process can be used:
Immerse in cold tap water for 5 minutes.
With a syringe, flush 3 times the lumen of
the part with 50 ml deionized water.
Place the instrument to a suitable rack and start
the automated cleaning.
The following validated (on Miele G 7735 CD)
automated cleaning process can be used:
2 min pre-washing with cold water.
Drain.
5 min cleaning with deionized water with
0.5% detergent of Neodisher MediClean
(Dr. Weigert, Hamburg) at 55°C.
Drain.
3 min rinsing and neutralization with
deionized water.
Drain
2 min final rinse with deionized water.
Also instructions of the manufacturer of the
washer disinfector shall be followed.
Automated
Automated
Automated Disinfection
Automated
Disinfection
Disinfection
Disinfection
Automated Thermal Disinfection in a disinfector
under consideration of national requirements in
regards to A0-Value accordingly to ISO 15883.
®
, AIRFLOW
®
and
®
, AIRFLOW
®
and
28/46