The following validated Pre-vacuum Moist Heat (steam) process can be used:
Parameters for the
Parameters for the
Parameters for the P P P P re re re re- - - - vacuum
Parameters for the
3 pre-vacuum phases
Sterilization temperature of 132°C
Pressure of 3 bar
Humidity of 100%
Holding time of 4 minutes minimum (full cycle)
Drying time of 20 minutes
It is the duty of the user to ensure that the reprocessing processes, including resources,
materials and personnel, are capable to reach the required results and maintained over the time:
keeping actual the validation of the reprocessing processes is under the responsibility of the user.
Storage
Store the sterilized instruments in a dry, clean and dust free environment at a temperature of
5°C to 40°C.
Service life
Refer to the section "Service life" of the "Technical Description" chapter for details.
The products have been designed for a large number of sterilization cycles. The materials used in
their manufacture were selected accordingly. However, with every renewed preparation for use,
thermal and chemical stresses will result in the ageing of the products.
If the number of permissible re-sterilization cycles is restricted, this will be stated in the
product's specific instructions (if any) and/or in the "Service life" section of the "Technical
Description" chapter.
Always replace products that present sign of worn-out or of early degradation, regardless of the
number of sterilization cycles left unused.
DO NOT expose the products to temperature exceeding the 138°C.
ENGLISH
vacuum Moist Heat
vacuum
vacuum
Moist Heat cycle:
Moist Heat
Moist Heat
FB-618/EN – Rev. A-01 - 2017.03.30