Preparation
Immediately after use, rinse the lumen(s) line of the part (device / instrument) and remove gross
soiling by wiping the instrument.
For AIRFLOW
®
and PERIOFLOW
for both lumens (water and powder) clearage before proceeding.
For PIEZON
®
:
Remove any installed instrument.
Remove the nose cap and separate aside the light guide and the red gasket.
For Endochuck file holder:
Remove any installed file.
Remove the screw and separate aside the small O-ring (gasket).
Safely transport to the reprocessing area to avoid any damage to the parts and contamination to
the environment and to the people involved in the reprocessing process.
Cleaning shall need to be performed within 1 hour from the use.
ENGLISH
®
: always carry out handpiece powder unclogging and check
FB-618/EN – Rev. A-01 - 2017.03.30