4.2.2 Connecting the VP cable to the transmitter
Standard connection for CO
Function
pH glass
Reference
Temperature 1
Temperature 2
Shielding
For further details check instruction manual of
the transmitter.
4.2.3 Calibation of pH electrode / transmitter 5100 e
Prepare electrode as described in section 4.1 "Start-
up". We recommend a 2-point calibration in pH buffer
pH 7.00 and pH 9.21. Before calibration, first remove
wetting cap. Normally you start with buffer pH 7.00 for
determining zero point and then buffer pH 9.21
for determining the slope. Please refer to instruction
manual of the transmitter for further details.
Important settings: Temperature sensor select Pt1000.
InPro 5000
52 002 465
transmitter 5100 e
2
VP cable wire
transparent
red
white
green
green/yellow
Min. immersion depth
25 mm / 1"
InPro 5000 CO
CO
Transmitter
2
Terminal connection
1
2
E
D
C
© 09 / 06 Mettler-Toledo AG
Sensor 12 mm
2
VP Cable
Printed in Switzerland