UTERINE POSITIONING
The RUMI Handle is always intended to be used with the Grip in the upward position
(see Figure 1).
TO ANTEVERT or insert into an anterverted uterus (see Figure 2): depress the Trigger and
turn the Grip clockwise lifting the uterus to achieve the desired degree of anteversion, up to
90°. Release the Trigger to lock the desired position.
TO RETROVERT or insert into a retroverted uterus (see Figure 3): depress the Trigger and
turn the Grip counterclockwise. A counterclockwise rotation of the Grip will result in the uterus
being lowered to a retroverted position down to 50°. Release the Trigger to lock the
desired position.
Figure 2
Antevert
RECOMMENDED CLEANING INSTRUCTIONS
Before cleaning the Handle, remove the Tip as per the RUMI
Directions for Use.
1. Prepare the enzymatic detergent according to the manufacturer's recommendations and
submerge the Handle in the detergent. Soak for ten (10) minutes.
2. Scrub the submerged Handle with a soft sponge and agitate. Clean lumens and crevices,
using a pipe cleaner/brush or flush with a syringe when appropriate. Actuate any moving
parts to loosen trapped soil.
3. Rinse in warm 100 °F - 120 °F (38 °C - 49 °C) tap water for one (1) minute. Thoroughly flush
all lumens and other difficult to reach areas. Actuate while rinsing.
4. Ultrasonically clean the Handle for ten (10) minutes in a neutral pH detergent (Neutrad
acceptable alternative). Prepare the detergent according to the manufacturer's
recommendations.
5. Rinse the Handle with clean tap water, actuating moving parts while rinsing, for one (1)
minute; repeat rinse twice.
6. Dry the Handle thoroughly with a clean, lint free cloth.
7. Repeat this cleaning procedure if the Handle appears to be soiled after initial cleaning and
until there is no sign of soil on the device.

RECOMMENDED STEAM AUTOCLAVE STERILIZATION PARAMETERS

• The instrument must be thoroughly cleaned of all foreign matter prior to sterilization.
• Follow the manufacturer's instructions for operation and loading of steam autoclave.
• There must be direct steam exposure to all surfaces of the instrument being sterilized.
Pre-vacuum Cycle:
Preconditioning Pulses: 4
Exposure Temperature: 270 °F - 273 °F (132 °C - 134 °C)
Exposure time: 4 minutes
(continued)
Figure 3
Retrovert
®
Uterine Manipulator Tips
®
or