Recommended Steam Autoclave Sterilization Parameters - CooperSurgical RUMI II Gebrauchsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 5
4. Close and lock the Latch, and ensure that the Manipulator
Handle is securely attached to the Adapter with no free
movement.
CAUTION: When securing the Latch, use care to avoid
contacting the patient or other accessories in the surgical field.
Removing the Manipulator Handle from the Adapter
To detach the Manipulator Handle, unlock and open the Latch
on the Adapter while holding the Manipulator Handle and
separate it from the Adapter.
Note: There is no need to depress the foot pedal and move the
UPS arm to separate the Manipulator Handle from the Adapter.
Removing the Adapter from the UPS
To detach the Adapter from the UPS Arm after surgery is
complete, press the Release Button on the Arm (Figure 3) and
pull the Adapter straight out without rotation.
RECOMMENDED CLEANING INSTRUCTIONS
1. Prepare the enzymatic detergent according to the
manufacturer's recommendations and submerge the Adapter
in the detergent. Soak for ten (10) minutes.
2. Scrub the submerged Adapter with a soft sponge or brush
and agitate. Clean crevices, using a pipe cleaner/brush or
flush with a syringe when appropriate. Actuate any moving
parts to loosen trapped soil.
3. Rinse in warm (100 °F - 120 °F, 38 °C - 49 °C) tap water for
one (1) minute. Thoroughly flush difficult to reach areas.
Actuate while rinsing.
4.
Place the Adapter into a bath containing warm (100 °F - 120 °F,
38 °C - 49 °C) water. Agitate the Adapter by hand for at least
one (1) minute. Repeat this process two (2) additional times.
5. Ultrasonically clean the Adapter for twenty (20) minutes in a
neutral pH detergent (Neutrad
Prepare the detergent according to the manufacturer's
recommendations.
6. Rinse the Adapter with clean, tap water, actuating moving
parts while rinsing, for one (1) minute; repeat rinse twice.
7. Dry the Adapter thoroughly with a clean, lint free cloth.
8. Repeat this cleaning procedure if the Adapter appears to be
soiled after initial cleaning and until there is no sign of soil
on the device.
RECOMMENDED STEAM AUTOCLAVE STERILIZATION
PARAMETERS
Prevacuum Cycle
Preconditioning Pulses: 4
Exposure temperature: 270 °F - 273 °F (132 °C - 134 °C)
Exposure time: 4 minutes
Gravity Displacement Cycles
Exposure temperature: 270 °F - 275 °F (132 °C - 135 °C)
Exposure time: 15 minutes
Exposure temperature: 250 °F - 253 °F (121 °C - 123 °C)
Exposure time: 40 minutes
Figure 3. UPS Arm
UPS Handle
Release Button
UPS Arm
®
or acceptable alternative).