Signal Description

Start/Stop: input signal to start or stop the pump. With the supplied
cover connector the START/STOP (+) signal is connected to the +24
Vdc pin and the START/STOP (-) signal to the GROUND pin: in this
condition the pump automatically starts as soon as the controller
recognises the input supply ("Plug & Pump").
Interlock: input signal to control the pump rotation. With the
supplied cover connector the interlock (+) signal is connected to the
+24 Vdc pin and the interlock (-) signal to the GROUND pin.
Speed setting: PWM input signal to set the pump speed. The PWM
signal characteristics must be the following:
Figure 14
The duty cycle percentage is referred to the low level portion of the PWM signal
NOTE
TV 1001 Navigator User Manual /
frequency: 100 Hz +/-20 %
amplitude: 24 V max
duty cycle range: from 25 % to 75 %
corresponding to an output frequency from 272 Hz to 633 Hz (see
the following diagram):