10. Technical specifications

Model No.
Measurement method
Measurement range
ECG range/sample rate
Display accuracy
Measurement inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible operating conditions
Permissible storage conditions
Power supply
Battery life
Classification
BM 95
Oscillometric, non-invasive blood pressure measurement on the
upper arm.
Single-channel ECG in freely selectable front positions / ECG sig-
nal ground (earth) referenced.
Cuff pressure 0 – 300 mmHg,
systolic 60 – 280 mmHg,
diastolic 30 – 200 mmHg,
Pulse 30 –180 beats/minute
0.05 to 40 Hz/256 Hz
Blood pressure: ± 3 mmHg or 2% of the value shown
Pulse: <± 5% of the value shown
Max. permissible standard deviation according to clinical testing:
Systolic 8 mmHg/diastolic 8 mmHg
2x 60 memory spaces
Main unit: L 128 mm x W 128 mm x H 40 mm
ECG stick: D 25 mm x H 125 mm
Main unit: Approximately 430 g (without batteries, with cuff)
ECG stick: Approximately 40 g
22 to 42 cm
+10 °C to +40 °C, 30-85 % relative humidity (non-condensing)
-20 °C to +50 °C, 10-85 % relative humidity (non-condensing)
4x AAA batteries
For approx. 300 measurements, depending on levels of blood
pressure and pump pressure
Internal supply, IPX0, no AP or APG, continuous operation
Blood pressure: Application part, type BF
ECG stick: Application part type CF