• Steam sterilization, steam sterilizer according to EN 13060 and/or EN 285 and validated
according to EN ISO 17665
Program steps
Process
Sterilization temperature
Sterilization time (holding time at
sterilization temperature)
Drying time
Proof of fundamental suitability of the instruments for effective steam sterilization was
provided by an independent, accredited test laboratory, test report no. 17315 dated
30.11.2015. Typical conditions in clinics and medical practices as well as the process
described above were taken into account.
a  Do not sterilize in hot air!
a  Do not sterilize in STERRAD
a  In case of potential contact with prions (CJD – risk of contamination), destroy the inst-
rument and do not use it again.
Storage / Transportation:
Store in a dry place. Protect from sunlight. Store and transport in secure containers /
packaging.
For returns, products must be cleaned, disinfected, and packed in sterile packaging.
Any change to the product or deviation from these instructions for use waives the liability
of Sutter Medizintechnik.
Changes reserved. Current version available at: www.sutter-med.de
®
STERRAD
is a trademark of Johnson & Johnson, Inc.
FIG A: components
FIG B: assembly and functional test
FIG C: disassembly
FIG D: hazard notice
Parameters
Fractionated vacuum (3x)
132 °C (max. 138 °C plus tolerance
according to EN ISO 17665)
Min. 3 minutes
Not specified
®
!