CH1:
LINE:
EXT:
(13)
„CAL/VAR"
To select the variable timebase uncalibrated or stepped calibrated.
(9)
Trigger LEVEL control
To select the trigger signal amplitude at which triggering occurs.
When rotated clockwise, the trigger point moves toward the positive
peak of the trigger signal. When this control is rotated counterclock-
wise, the trigger point moves toward the negative peak of the trigger
signal.
(8)
Trigger Slope switch
To select the positive or negative slope of the trigger signal (on
(on LEVEL control)
LEVEL control) for initiating sweep. Pulled in, the switch selects the
positive (+) slope. When pushed, this switch selects the negative (-)
slope.
(19)
EXT TRIG IN connector
For applying external trigger signal to the signal circuits.
CAUTION
To avoid damage to the oscilloscope do not apply more than 400 V (DC + Peak AC)
between "EXT Trig In" terminal and ground.
(31)
EXT Blanking Input
For applying signal to intensity modulate the CRT, trace brightness is
connector
reduced with a positive signal, and increased with a negative signal.
(17)
Probe Adjust
Provides a fast rise square wave of precise amplitude for probe
adjustment and vertical amplifier calibration.
(28)
Ground Connector
Provides and attachment point for a separate ground lead.
2-2. Basic Operating Procedures
2-2-1. Preliminary Control Settings and Adjustments
a) Before placing the instrument in use, set up and check the instrument as follows:
POWER switch (16)
INTEN control (2)
FOCUS control (1)
AC/GND/DC switch (25),(21)
VOLTS/DIV siwtch (26),(23)
X5 MAG switch (3)
Vertical POSITION controls (4),(7)
INV switch (6)
VARIABLE controls (27),(20)
V MODE switch (5)
TIME/DIV switch (15)
34
In case of vertical mode switch is CH1 which automati-
cally becomes registry source.
In case of vertical mode switch is CH2 which automati-
cally becomes registry source.
When Vert Mode Switch is dual, the signals are dis-
played with ALT Mode in all TIME/DIV range.
When there is the signal in the CH1, you may select trig-
ger source CH1.
Position selects a trigger derived from the AC poser line,
this permits the scope to stabilize display line, related
components of a signal even if they are very small com-
pared to other components of the signal.
Position selects the signal applied to the EXT TRIG IN
connector.
OFF (relaesed)
Mid rotation
Mid rotation
DC
10mV
X1
Mid rotation
Norm
Fully CW
CH1
1ms
Abb. 2 – 6 Spitze - Spitze - Spannungsermittlung
Abb. 2 – 7 Gleichspannungsmessung
B Gleichspannungsmessung
Um eine Gleichspannung zu ermitteln, gehen Sie wie folgt vor:
1. Stellen Sie das Oszilloskop gemäß 2 – 2 - 1 ein.
2. Stellen Sie den AC/GND/DC – Schalter des Kanals, an welchem das Signal anliegt, auf GND. (25) oder
(21).
3. Stellen Sie den Strahl mit dem betreffenden Positionssteller (4) oder (7) auf die Mitte des
sters ein (= die waagerechte skalierte Mittellinie). Dieser POS – Steller darf nun bis zum Ende der Mes-
sung nicht mehr verstellt werden.
4. Stellen Sie nun den AC/GND/DC – Schalter (25) oder (21) auf DC, aber beachten Sie unbedingt die maxi-
malen Eingangsgrößen. Alle Signale oberhalb der Mittellinie sind nun positiv, darunter negativ.
Achtung!
Vergewissern Sie sich, daß die Wechselgröße nicht auf eine sehr große Gleichgröße auf-
gelagert ist. Beachten Sie unbedingt die max. zulässigen Eingangsgrößen.
5. Ermitteln Sie den Abstand zwischen der positiven oder der negativen Linie und der Mittellinie durch
zählen der Skalierungsstriche bzw. der ,,Kästchen" = DIVISIONS und multiplizieren Sie diese mit der Ein-
stellung des VOLTS/DIV – Schalters,
Bildschirmra-
19