Herunterladen Inhalt Inhalt Diese Seite drucken

Operation; Switch On Device; Operating The Three-Way Cock; Overview: Controls Control Panel - Nouvag MD20 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
NOUVAG AG

6.1 Switch on device

The control device can be switched on or off with the main switch "I/O" (at the back!).

6.2 Operating the three-way cock

With the three-way cock four different positions can be
selected for cooling:
Position c: motor 1 connected to cooling
Position d: both motors connected to cooling
o This position should be avoided,
because otherwise liquid may also leak out
if motor is not used!
Position e: motor 2 connected to cooling
Position f: no cooling!

6.3 Overview: controls control panel

A
B
C
D
A) Display:
Display of various operational parameters.
B) Keys „ Program 1" to „ Program 3" :
Choose Program 1 to 3 (choose for both motors separate)
C) Key „ Print" :
Print out of operating data (functions only with connected printer!)
D) Key „ Motor 1/2" :
Change between the connected motors.
E) Option keys „ Hand piece" :
„-" and „+": select reduction / transmission or fixed setting.
F) Option keys „ Speed" :
„-": reduce rotational speed
G) Option keys „ Torque" :
„-": reduce torque
H) Option keys „ Pump" :
„-": reduce pump speed
No. 31842 14/06

6 Operation

„+": increase rotational speed
„+": increase torque
„+": increase pump speed
e
c
d
E
F
G
H
MD20
f
7

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis