1. Specification
Combustion air temperature (TL):
Reading:
Measuring principle:
semiconductor resistor (Si+PTC)Measuring range:
-19.9 to 99.9°C
Accuracy:
according to "BlmSchV" requirements
1.2.
Calculated values:
Waste gas loss in % as a function of the fuel according to "BlmSchV" (QA)
ETA: efficiency from 0 to 120 %
CO
: carbon dioxide - content in %
2
Air surplus number Lamda (Lam) (e.g. 1.25 for 25 % of air surplus)
Dew point in °C (TP)
Amount of condensate (Kon) in kg/m
1.3.

Extra functions in the basic version

Differential pressure recording possible via external pressure connection (upstream
of the charging terminal) for adjusting the flow or jet pressure. Measuring range: -
100 to +4000 Pa
Display of battery condition
Data transfer to a thermal printer via infra-red interface
Data memory of 10 datarecords, as well as a permanent datalogging (if equipped
with infra-red interface)
1.4.

Technical Data:

Power supply:
Current input:
Voltage:
Storage temperature:
Operating temperature:
Dimensions:
1.5.
Gas intake and preparation:
The waste gas will be sucked in via the probe by means of a membrane pump and
carried to the condensate trap. That is where a heat exchanger will cool the waste
gas. A cotton filter will retain the dust particles. Lastly, a water stop filter (PFTE
membrane) will keep the membrane pump and the measuring cells free from any
water and dust.
Manual Instruction
°C
3
4 Mignon batteries or 4 accus possible
battery status on display
approx. 60 mA with pump operating
4.4 to 6.0 volt
-20 to +50°C
+10 to +40°C if accuracy levels indicated are to be
observed
115 x 84 x 55 mm (without probe)