operating instructions

Connecting The iFill Oxygen Cylinder To The DeVilbiss iFill
Personal Oxygen Station
Danger
Fire Hazard. Do not lubricate. Do not allow grease or oil from
your hands or other source to come into contact with the
regulator or cylinder valve connection. These solutions may
be flammable and cause injury.
Warning
Do not modify ANY connections on the personal oxygen
station. NEVER use tools of any kind to connect/disconnect
the cylinder and the oxygen station. Severe injury and/or
damage may occur.
Do not drop oxygen cylinders. Use two (2) hands when
handling/transporting oxygen cylinders. Otherwise, injury or
damage may occur.
– Refer to the General Dangers & Warnings and the
NOTE
Handling Warnings in this manual.
1 .
Plug the power cord into the wall outlet .
2 .
Set the rotary selector on the cylinder to "OFF ."
3.
Remove the oxygen fill connector cover from the fill
connector .
4 .
Position the cylinder over the cradle while aligning the
nipple connector on the oxygen cylinder with the fill
connector . Press until cylinder "clicks" into place . NOTE–
When using the smaller M4 or M6 size cylinders, it is
necessary to place the oxygen fill connector cover/cylinder
adapter in the cradle to help support the cylinder . When
using the larger ML6, C, D, or E cylinders, the oxygen fill
connector cover/cylinder adapter is not necessary and can
be allowed to hang over the side .
filling the ifill oxygen cylinder
– Do not use cylinder while filling.
NOTE
– Refer to the General Dangers & Warnings and the
NOTE
Handling Warnings in this manual.
1 .
Make sure the personal oxygen station is plugged in .
2 .
Make sure the cylinder is securely connected to the oxygen
station . Refer to Connecting The iFill Oxygen Cylinder To
The DeVilbiss iFill Personal Oxygen Station .
3 .
Push the start/stop switch on the control panel .
4 .
The following sequence of events should occur:
a . The FILLING (green) light will illuminate while the
cylinder is filling (refer to iFill Oxygen Cylinder Fill
Times) .
b . The FULL (green) light will illuminate when the cylinder
is full . Proceed to Removing the iFill Oxygen Cylinder .
– If the Service (RED) light and audible alert are on,
NOTE
proceed to Trouble shooting in this service manual.
removing the ifill oxygen cylinder
Warning
Do not modify any connections on the personal oxygen
station. Never lubricate the connections. Never use tools of
any kind to connect/disconnect the oxygen cylinder and the
LT-1929
U N P A C K I N G A N D S E T U P
oxygen station. Otherwise, severe injury and/or damage may
occur.
Use extreme care when handling and filling an oxygen
cylinder. Full oxygen cylin ders are under pressure and can
become a projectile if dropped or mishandled.
NOTE–
Handling Warnings in this manual.
1.
2.
3.
4 .
NOTE
whenever cylinders are not being filled.
setting the ifill oxygen cylinder rotary
selector to the prescribed setting
1.
2 .
Warning
Changing the L/min. setting on the rotary selector will affect
the dose of oxygen delivered, DO NOT readjust the L/min.
setting unless directed by your physi cian.
3.

inDicator LigHt expLanation

None
Standby
(green)
Filling
(green)
Full (green)
Service (red)
and Audible
Alert
6
Refer to the General Dangers & Warnings and the
Remove the full oxygen cylinder by pushing down on the fill
port collar located under the cylinder nipple connector while
using your other hand to steady the cylinder .
Lift up on the cylinder to remove from the oxygen fill
connector . The green Standby light will turn on after a short
delay .
Oxygen will escape from the fill connector for a short time
after the cylinder has been removed . This may be
accompanied by a "pop" with a rush of air . This is normal .
Perform one (1) of the following steps:
a.
If desired, fill another oxygen cylinder. Refer to the
DeVilbiss iFill Personal Oxygen Station Operation
Checklist .
b.
Place the oxygen fill connector cover onto the fill
connector .
– The oxygen fill connector cover should always be used
Attach the nasal cannula to the cannula fitting of the iFill
oxygen cylinder . NOTE– If it is an international cylinder,
open the cylinder valve first.
Turn the rotary selector to the L/min . setting prescribed by
your physician .
To turn the flow of oxygen off, turn the rotary selector
counterclockwise to the "OFF" position .
Indicator
Light Color
Oxygen Station Status
Oxygen station is
unplugged .
Oxygen station is ready to
begin filling cylinders .
Oxygen station is on and
cylinder is filling .
Cylinder is full .
Oxygen station is on but
cylinder is not filling due
to internal failure .
Effect
None
Cylinder filling
not started
Cylinder is filling
Remove
cylinder
Service is
required .

Werbung