introDuction

This service manual was designed to provide DeVilbiss Healthcare qualified service technicians and homecare providers with the
proper maintenance, service, safety, and repair procedures for the DeVilbiss iFill Personal Oxygen Station .
Read and understand all the information contained in this service manual before attempting to operate or perform any maintenance
on the oxygen station .
iFill is a stand-alone oxygen cylinder filling station that produces highly concentrated oxygen for therapeutic applications. Because
of the stand-alone feature, iFill can be used in conjunction with any concentrator .
– DeVilbiss reserves the right to alter or change the design of the DeVilbiss iFill Personal Oxygen Station series. Hence, slight
NOTE
differences in construction or components may exist between the unit in hand and what is described in this manual.

iec sYmBoLs

Attention, consult instruction guide
AC current
Electric S hock Hazard . Do Not Open
This device contains electrical and/or electronic equipment that must be recycled per EU Directive 2012/19/EU-Waste
Electrical & Electronic Equipment

important safeguarDs

When using electrical products, especially when children are present, basic safety precautions should always be followed . Read all
instructions before using . Important information is highlighted by these terms:
Urgent safety information for hazards that will cause serious injury or death.
Danger
Important safety information for hazards that might cause serious injury.
Warning
CAUTION
Information for preventing damage to the product.
NOTE
Information to which you should pay special attention.
Important safeguards are indicated throughout this service manual; pay special attention to all safety information .
reaD aLL instructions Before serVicing.
generaL Dangers & Warnings
In order to ensure the safe installation, assembly, and operation of the personal oxygen station, these instructions must be followed:
Danger-no smoking
• DANGER-Electric Shock Hazard. If service is required, it should be done by a DeVilbiss Healthcare provider or authorized
service center.
• DANGER-Fire Hazard. Do not lubricate. Do not allow grease or oil from your hands or other source to come into contact with
the regulator or cylinder valve connection. These solutions may be flammable and cause injury.
• DANGER-Keep the personal oxygen station at least 5 feet (1.6 m) from hot, sparking objects or naked sources of flame.
Position your unit at least 6 inches (16 cm) from walls, draperies, or any other object that might prevent the proper flow of
air in and out of your oxygen station. The oxygen station should be located so as to avoid pollutants or fumes.
• DANGER-Oxygen causes rapid burning. Do not smoke while your personal oxygen station is operating, or when you are
near a person utilizing oxygen therapy. Keep the oxygen station at least 5 feet (1.6 m) from hot, sparking objects or naked
sources of flame.
• DANGER-Do not lay the cannula down while the cylinder is delivering oxygen. High concentrations of oxygen can cause
rapid burning.
G E N E R A L I N F O R M A T I O N
Type BF applied part
Double Insulated
Danger-No Smoking
3
LT-1929