Set Up & Operating Instructions
Refer to the instruction guide (A-306-1 or A-306-2) for set up
and operating instructions. An overview is below:
CAUTION–The plug on the DeVilbiss iGo is a three-prong plug that
must be connected to a grounded receptacle.
CAUTION–Replace the power cord if it is damaged.
NOTE–Every time a button is pressed or a change to the power
source occurs, the iGo gives a short audible beep.
Press the Power button and hold for one second to turn iGo ON.
All the lights on the control panel illuminate briefly and an audible
alert sounds momentarily. Some lights will remain illuminated;
others will not.
When using AC Power:
In PulseDose mode: After a few seconds, the External Power,
Normal Oxygen, and Flow Rate lights will remain lit. The Flow
Indicator light pulses ON with each breath. If a battery is installed,
the Battery Status lights illuminate to indicate battery charge level
or cycle to indicate battery is charging.
In Continuous Flow mode: After a few seconds, the External
Power, Normal Oxygen, Flow Indicator, and Flow Rate lights will
remain lit. If a battery is installed, the Battery Status lights
illuminate to indicate battery charge level or cycle to indicate
battery is charging.
When using External DC (car adapter)
Power:
DO NOT leave the iGo or DC Adapter plugged into the vehicle
without the engine running or attempt to start the vehicle
while the DC Adapter is connected . This may drain the
vehicle's battery.
In PulseDose mode: After a few seconds, the External Power,
Normal Oxygen, and Flow Rate lights will remain lit. The Flow
Indicator light pulses ON with each breath. If a battery is installed,
the Battery Status lights illuminate to indicate battery charge level.
The battery will not charge while operating on External DC power.
In Continuous Flow mode: After a few seconds, the External
Power, Normal Oxygen, Flow Indicator, and Flow Rate lights will
remain lit. If a battery is installed, the Battery Status lights
illuminate to indicate battery charge level. The battery will not
charge while operating on External DC power.
U N PA C K I N G A N D S E T U P
5
When using Internal DC (battery) Power:
In PulseDose mode: After a few seconds, the Normal Oxygen and
Flow Rate lights will remain lit. The Battery Status lights illuminate
to indicate battery charge level. The Flow Indicator light pulses ON
with each breath.
In Continuous Flow mode: After a few seconds, the Normal
Oxygen, Flow Rate, and Flow Indicator lights will remain lit. The
Battery Status lights illuminate to indicate battery charge level.
NOTES:
When turned on, the iGo will START at the last mode and flow
setting used.
Continuous Flow mode provides a continuous supply of
oxygen at a specified flow level.
PulseDose mode provides a bolus of oxygen at a
specified flow level when a breath is detected. An alert
will beep after 30 seconds if a breath is not detected. If
30 additional seconds elapse with no breath detected,
the unit automatically switches to continuous flow mode
at the last continuous flow setting used.
PulseDose mode increases the amount of operating time
when using battery power.
PulseDose delivers oxygen in a very short puff. It does not
deliver oxygen continuously. The length of time that
PulseDose delivers oxygen will not vary from breath to breath.
The time is set in correlation to the PulseDose setting.
PulseDose is designed to prevent the delivery of pulses more
than every 2 seconds. If the breath rate is greater than the
maximum breathing rate, this feature prevents delivery of
excessive oxygen by not dosing on every breath.
Verify that the flow rate is on the correct prescription
setting or use the Flow Rate Setting buttons to adjust the
flow rate
NOTE–You may lock the flow so that it cannot be changed from the
prescribed setting: First set the continuous flow rate to the
prescribed setting, then set the PulseDose flow rate. Turn the unit
OFF and while OFF, press and hold the Mode Select button for 5-6
seconds. A short beep will sound when the lock is complete. Verify
that the settings are locked. To unlock the settings: With the unit
OFF, press and hold the Mode Select button for 5 to 6 seconds.
The unit will beep when the unlocking takes affect. Verify that the
lock is inactive.