HYDRIM M2 G4 Installation Instructions
Checking Water Quality
The quality of the water being used in the HYDRIM to clean the medical instruments is very
critical to achieving satisfying cleaning results and to protecting the instruments and the
internal parts of the HYDRIM from deterioration. Dissolved solids in the water can cause
stains, spots, and corrosion on instruments and the internal parts of the HYDRIM.
Before installing and using the HYDRIM, SciCan recommends testing the water and
recording the results for water hardness, water conductivity, and pH-value for future
references here.
The following table lists the desired levels of the water parameters. If the test results of
any of the water parameters exceed the stated problem levels, a full water analysis is
recommended and the installation of a water treatment system may be required to achieve
satisfying cleaning results with the HYDRIM.
Water Parameter
pH value
Hardness
Water conductivity
The HYDRIM is equipped with an air gap / anti-suction device to prevent backflow of dirty
water into the water supply.
Water hardness
• Water conductivity
• pH value
Action required (if any):
Desired
Level
7.0 – 7.5
< 85
< 274
• Hot water
• Cold water
Units
S.U.
mg/l
microS/cm
Page 10
Problem
Level
< 6.8 or > 8.5
> 100
> 469