Error
Possible cause
Measured values/
Too little rest before
pulse values are
taking the measure-
extremely high
ment. Movement or
or low.
talking during the
measurement.
Incorrect measure-
ment posture
Measurements
Blood pressure is not
taken in direct
a fixed value. It can go
succession
up or down by more
produce different
than 20 mmHg within
measured values.
a few minutes.
Measurement
Device detects a fault
was interrupted,
or that the diastolic
the cuff deflated
value is too low.
to 0 and the
Movement during
measurement
measurement
restarted.
Dashes in display
No measured values
after pressing
have been stored.
Memory button
Batteries flat
The measurement
is interrupted and
the
symbol
appears.
The
symbol
Low batteries
appears during
measurement but
the measurement
continues.
No display after
Batteries inserted
device is switched
incorrectly
on
Batteries flat
Battery contacts dirty.
Date and time are
Function was uninten-
not visible in the
tionally deactivated
display
EN-56
Technical information
Remedy
Repeat the measurement
after a break of 3-5 minutes.
Pay attention to the usage
information on page 42.
Repeat the measurement
and maintain the correct
posture. (Page 45)
In order to obtain compara-
ble results, always carry out
the measurements under
the same conditions.
Device shows no fault.
Repeat measurement
If necessary, interrupt mea-
surement and repeat after 5
minutes' rest.
Repeat measurement
Replace batteries.
You can still perform
approx. 30 measurements,
but should keep spare
batteries available.
Check position of batteries.
Replace batteries.
Clean battery contacts with
a dry cloth.
Activate the date and time
function as described on
page 43.
Technical information

2. Customer service

Device repairs may only be performed by the manufacturer or an
expressly authorised body. Please contact:
UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Germany
Tel.: +49 (0) 9342/924040
Fax.: +49 (0) 9342/924080
Email: info@uebe.com
Website: www.uebe.com

3. Technical data

Model:
visomat
®
Size:
88 mm x 64 mm x 26.4 mm
Weight:
approx. 101 g without batteries and cuff
Display:
LCD display (liquid crystal display) for measured
values and check displays
Memory:
2 memories each capable of storing 120
readings
Measurement me-
Oscillometric determination of systole, diastole
thod:
and pulse
Reference method of
Auscultatory measurement
the clinical test:
Pressure display
0-300 mmHg
range:
Measurement range:
Systolic: 50-250 mmHg
Diastolic: 40-150 mmHg
Pulse measurement: 30-199 pulses/minute
Measurement pre-
Blood pressure measurement (cuff pressure):
cision:
± 3 mmHg
Pulse rate: ± 5%
Power supply:
2 x 1.5 V AAA LR03 alkaline manganese batte-
ries
Power consumption:
2 W
handy express
EN-57