Procedure:
Warning!
Take care that no dust enters the transmitter.
1. Transmitter: remove cover (a, Fig. 4) from the connection compartment.
2. Remove blank cover for the cable entries for channel 1 and channel 2.
3. Remove special cable entry (supplied with sensors). Run both sensor connecting
cables through the cover (b) of the cable gland and into the connection compart-
ment.
4. Position the cable fixing sleeves (c) of the sensor cables exactly next each other
(Detail C). Position the ground contact terminals (d) and screw them tight.
f
e
b
Fig. 5:
Connecting the sensor connecting cable
5. Spread the rubber seal (e) along the side slit using a suitable tool (e.g. a large screw-
driver) so that both sensor cables are clamped within. Push the rubber seal up into
the cable gland bracket (f). Close the cover of the cable gland (b) so that it is tight.
6. Plug in the sensor cable connectors analogue to the arrangement in Fig. 5.
7. Transmitter: secure cover (a) on the connection compartment.
g
d
<
c
C
13