Note
Check to make sure that the sensors are a matched set with the same serial numbers and
marked with an "A" and "B" (e.g., 19256A and 19256B).
Proceed to back to Sensor Installation (Page 10) if necessary.
4.3
Commissioning
1. From the [Meter Type] menu press the <Right Arrow> to select [Single Channel] and then
2. The [Channel Setup] menu will appear.
3. Scroll down to [Install Complete]. Press the <Right Arrow> key.
4. Press the <Down Arrow> key and select [Install].
5. Press <ENTER> key. Flowmeter will go through drives.
6. Observe the Measured Vs window and verify a correct sound velocity measurement (if
7. Press the <Down Arrow> to accept sound velocity value.
8. The flowmeter is now ready to report flow.
See also
Refer to I/O Connection and Wiring (Page 33) for input/output wiring and data spanning
procedures.
FST020 IP65 NEMA 4X
Operating Instructions, 07/2010, A5E03086489-01
press <ENTER>.
known).
①
Select [Install].
Figure 4-2
Measuring Flow
Commissioning
4.3 Commissioning
27