Installing the Sensor
1. Apply a continuous 1/8-inch bead of couplant compound across the center (the long way)
2. Place the sensor on the pipe center in the middle of one of the areas you have cleaned.
FST020 IP65 NEMA 4X
Operating Instructions, 07/2010, A5E03086489-01
of the sensor emitting surface.
Coupling Compound
Emitting Surface
Back End
Figure 2-4
Sensor
Front View
Spring Clip (Not present on some models)
Sensor Clamping Screw
Spacer Bar
7ME39600M Mounting Frame
Figure 2-5
Sensor Installation
Make sure its cable connector is pointing away from the other cleaned area.
2.3 Installing the Sensor
Sensor
Angled Edge
F-Connector
Pipe
Mounting Strap Note: Optional 2nd
Mounting Strap shown. Larger pipes,
over 30 inches may need additional
support.
Spacer Bar Platform and Clamping Screw
Spacer Bar Reference Hole
Orientation of sensor at 9 o'clock position
Installing/Mounting