Installing/Mounting
2.3 Installing the Sensor
3. Holding the sensor in place, slide the mounting strap over it and align the Strap Guide pin
4. While tightening strap, check to ensure that the front label is centered on the pipe and that
5. Repeat Step 2 with the second sensor leaving it still attached to the Spacer Bar. Apply a
6. Place the sensor in the middle of the second prepared area. Secure the Spacing Bar to the
7. Put mounting strap and Strap Guide around sensor and hand tighten. Sight along Spacer
14
with the mounting hole in the sensor. Tighten mounting strap with a screwdriver.
there is equal space on both edges. Also, make sure to position strap for easy access to
the mounting strap adjusting screw.
Sensor
Label
Front View
Figure 2-6
Sensor Alignment
continuous 1/8-inch bead of couplant compound across the center (the long way) of the
sensor emitting surface.
already mounted sensor by inserting the sensor index screw through the REF hole on the
Spacer Bar.
Bar to ensure axial alignment to the pipe. Adjust if necessary but do not overtighten. Ensure
that the sensors do not move while tightening.
FST020 IP65 NEMA 4X
Operating Instructions, 07/2010, A5E03086489-01