• Do not use detergents or solvents.
• On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.

9. Rectifying faults

In case of faults, the
_ message appears in the display.
Error messages may appear if:
• systolic or diastolic pressure could not be measured (
appears on the display),
• systolic or diastolic pressure was outside the measurement
range ( or Lo appears on the display),
• the cuff is fastened too tightly or loosely (
on the display),
• the pump pressure is higher than 300 mmHg (
on the display),
• pumping up takes longer than 160 seconds (
the display),
• there is a system or device error (
pears on the display),
• the batteries are almost empty
• the data could not be sent to the PC (
display).
In such cases, repeat the measurement. Ensure that the cuff
line is correctly attached and that you do not move or speak. If
necessary, reinsert or replace the batteries.
or
or
appears
appears
appears on
,
,
or
ap-
,
appears in the

10. Specifications

Model no.
Type
Measurement
method
Measurement range
Display accuracy
Measurement
inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible operating
conditions
Permissible storage
and transport condi-
tions
Power supply
Battery life
26
BM 45
M1502
Oscillometric, non-invasive blood pres-
sure measurement on the upper arm
Cuff pressure 0 – 300 mmHg,
systolic 60 – 260 mmHg,
diastolic 40 –199 mmHg,
Pulse 30 –180 beats/minute
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
L 165 mm x W 107 mm x H 50 mm
Approx. 284 g (without batteries)
22 to 36 cm
+10 °C to + 40 °C, ≤ 90 % relative air
humidity (non-condensing)
- 20 °C to + 55 °C, ≤ 90 % relative air hu-
midity, 800 –1050 hPa ambient pressure
4 x 1,5 V
AA batteries
For approx. 250 measurements, de-
pending on the blood pressure level
and/or pump pressure