Herunterladen Inhalt Inhalt Diese Seite drucken

Novasina AW SPRINT Bedienungsanleitung Seite 46

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

01:Print offset
0.000aw
01:Beep durat.
010 Seconds
46
Recommendation: For fast access to a stable value we recom-
mend selecting the value "1" and continuing to observe the change
in the aw value under "Actual" after the stable value has been
displayed. If the stable value deviates from the actual limit value by
more than desired, the humidity stability factor may be increased.
– Setting Print Offset
You set the so-called Print Offset (factory setting: 0.000aw) in this
submenu point.
If, on repeated measurements of the same product, it is established
that the stable value always deviates by the same amount from the
actual limit value, a correction value (the so-called "Print Offset")
may be established for the protocol printout. The Print Offset is only
indicated on the "Stable" protocol and does not influence the stable
value shown in the display.
– Setting the length of the stable signal tone
In this submenu point you set how long the signal (beeper) is to
sound after the stable value has been reached.
– Factory setting:
10 seconds
– Setting range :
Beep off (no signal) ... 600 seconds

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis