3. Specifications

Electrical data

Power supply U
B
Influence of power
supply
Permissible residual
ripple
Output signal
Permissible
max. load R
A
(variant 1 - 3)
Effect of load
(variant 1 - 3)
Electrical zero point
Long-term stability
of electronics
Electr. output signal
Linear error
Electrical connection
Wiring protection
As a power supply, an unstabilised DC power supply is sufficient in the range of
the specified limits. It is important to ensure that the applied power supply is at
least greater than the maximum required voltage drop from the external display
and evaluation units; i.e. the voltage at the instrument must not fall below 10 V.
WIKA operating instructions differential pressure gauge, model DPGT40
DC V
12 < U
14 < U
15 < U
% FS/10 V
≤ 0.1
% ss
≤ 10
Variant 1
4 ... 20 mA, 2-wire, passive, per NAMUR NE 43
Variant 2
4 ... 20 mA, 2-wire, Ex version
Variant 3
0 ... 20 mA, 3-wire
Variant 4
0 ... 10 V, 3-wire
R
≤ (U
A
however max. 600 Ω
% FS
≤ 0.1
Through a jumper across terminals 5 and 6
% FS/a
< 0.3
≤ 2.5 % of measuring span (option ≤ 1.6 %)
% of span
≤ 2.5 % (option ≤ 1.6 %), terminal method
Via angular connector, 180° rotatable,
wire protection, cable gland M20 x 1.5,
incl. strain relief, connection cable: Outer
diameter 7 ... 13 mm, conductor cross-
section 0.14 ... 1.5 mm
up to 70 °C
Angular connector: IP65 per EN/IEC 60529
≤ 30 (Variant 1 + 3)
B
≤ 30 (Variant 2)
B
≤ 30 (Variant 4)
B
- 12 V)/0.02 A with R
B
, temperature resistance
2
in Ω and U
in V,
A
B
9
EN

Werbung