Herunterladen Diese Seite drucken

Ace instruments ALP-1 med Bedienungsanleitung Seite 9

Vorschau ausblenden Andere Handbücher für ALP-1 med:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

ACE ALP-1 med Breathalyzer
Operating Manual
Author's Liability:
All information herein was created conscientiously and under the assumption
of correctness. Hence, the author cannot be held liable for the use of any kind
whatsoever and/or any ensuing consequences. Furthermore, the author is not
liable for any costs or expenses for damage that result from any information in
this document. Any information is subject to change without prior notice.
Introduction:
Thank you for purchasing our ACE ALP-1 med Breathalyzer.
This breathalyzer complies with the Medical Device Directive 93/42/EEC and the
DIN Norm EN 15964:2011. This device measures alcohol concentration in human
breath using an electrochemical sensor and calculates the blood alcohol level.
The test results are very reliable and have a tolerance of +/- 0.05‰ (per mill) with
reference to 1.00‰. The device must only be used pursuant to the instructions
mentioned in this manual. The measurement ‰ (per mill) is the default setting
for the usage in the EU.
Short description:
The ACE ALP-1 med Breathalyzer is an easy-to-use alcohol tester with an
electrochemical sensor. The test results can be saved in the device or printed
via a cable or IrDA connection on the mobile ACE printer (optional available).
It can be used in a normal mode (with mouthpieces) and in a fast mode (with the
delivered breath funnel). In the normal mode the test result is shown on the display
with 3 digits (2 decimals*) with measurement setting „per mill" and it can be inter-
preted 1:1 – so if the shown test resultis, e.g., 0.36 then this means 0.36‰ = per mill
(for UK: if the shown test result is, e.g., 0.18, then this means 0.18mg/L, for USA: if the
shown test result is, e.g., 0.036, then this means 0.036% BAC).
In the fast mode you only see „PASS" on the display (= your alcohol level is below
the Individually defined limit) or „FAIL" (= your alcohol level is above the limit). The
default setting = „PASS" up to 0.50‰ and „FAIL" above 0.50‰.
The fuel cell sensor inside the ACE ALP-1 med Breathalyzer can be calibrated
and as alternative the sensor can also be exchanged against a new one.
For a calibration service you can send the device or the exchanged sensor to
ACE Instruments (see www.ace-instruments.net). For each breath test in normal
mode please use a new mouthpiece as rests of alcohol inside the used mouth-
piece can negatively influence another measurement with this used mouthpiece.
*4 digits and 3 decimals if the measurement setting is % BAC.
Scope of delivery:
1 Digital fuel cell ACE ALP-1 med Breathalyzer
2 AA batteries (1.5 V)
5 Mouthpieces & 1 Breath funnel (for tests in fast mode)
1 MAX30 CD (PC Software)
1 USB cable (for the connection with a PC)
1 Leather bag
1 Hand strap
1 Quick Start Guide
1 Operating Manual
1 Medical certificate
Optional equipment: ACE printer for ACE ALP-1 med

Werbung

loading