Administer menu:
To get into the administration area please turn on the device by pressing the
power button and afterwards press the button "m" on the front of the device
constantly for 2-3 seconds. Press the power button again and then type in the
pin code (default: 0530), to connect with a printer type in pin code "1000".
To change between the single menus please use the buttons Up/Down on
the front of the device) and press the power button to confirm your choice.
A. Contents of administer menu:
I. Information
II. General Setting
III. Measurement Setting
IV. Calibration
B. Details of chooseable contens in the administer menu:
I. Information:
a.
Last tests
b.
Last maintenance
c.
Total test count
d.
Date & Time
e.
Configuration
f.
Device info
II. General setting
a.
Set language (changes can only be done concerning the text which is shown
on the display before, during and at the end of a test operation; you can choose
English, German, Italian, Spanish and Russian)
b.
Set date
c.
Set time
d.
Printer (printer settings, please check that the device is with the printer cable
before you choose the connection "cable". A bluetooth connection is optional and
not available with this model. You can buy an optional mobile printer, with which
you can print "last data" (the data of your last measurement) and "selected data"
(the data which before have been defined in the menu under "range").
You can print "last data" also via an IrDA connection, therefore you have to hold
the IrDA-window on the front of the ACE ALP-1 med near to the IrDA-window of the
printer, then press the power button on the ACE ALP-1 med and the print of "last
data" will start.
e.
Save to SD card (not available with this model)
f.
Change PIN
g.
Bluetooth (not available with this model)
III. Measurement set.
a.
Measurement unit
Passive mode (on/off)
b.
c.
Limits level (set the measurement limits, this influences the warning tone + lightning)
d.
Blow criteria (setting of the minimum breathing volume
and minimum blowing time)
e.
Sampling type (Setting of 2 possible types of sampling:
1. Fixed volume = sampling is finished when the minimum volume is reached
2. End of breath = sampling is finished at the end of the exhalation)
f.
Backlight (on/off)
g.
Test count option (Since last CAL = a number of tests after the last calibration;
Lifetime total = total number of ever with the device done tests; OFF = function off)
h.
CAL required No. (setting of the calibration period)
i.
Tachograph (connection to the tachometer of a vehicle, this function is optional
and not chooseable with this model)
j.
Measurement mode (Normal or Fast), the default setting is Normal mode. To get
exact test results you always should use "Normal" mode. In the "Fast" mode tests
are done by using the bretah funnel which is in the scope of delivery and fixed
on top of the device. Be aware that in the „Fast" mode the test result is more
imprecisely and not shown as number on the display but only with the letter
"P" (Pass) and "F" (Fail).
IV. Calibration (this menu must only be used by the ACE customer service!)
a.
Last maintenance
Accuracy CHK date (date/time, the device saves max. values)
b.
c.
Calibration (for service tasks only)
d.
Accuracy check (check of accuracy, for service tasks only)
e.
Pressure Cal. (for service tasks only)
f.
Set last CAL date (for service tasks only)
g.
Set last ACC date (for service tasks only)
Calibration:
To maintain the accuracy of the breathalyzer, ACE recommends a calibration after
6 months (see reminder on the display). For a calibration service you can send the
device or the exchangeable sensor to ACE Instruments (see www.ace-instruments.
net). The calibration is not mandatory, but the accuracy of the breathalyzer will
decrease steadily without it.
We recommend sending it as a parcel, so that in cases of loss an inquiry or proof
of delivery is possible.
Due to changes in the product, the procedures stated can change. This can take
place without prior notice or adaptation of the operating manual and is no cause
for complaint, return or price reduction. The user is responsible for recognizing