Label of individual components:
LED light
1.
2.
IrDA window
3.
Digital LC display
4.
Up button
5.
Menu / down button
6.
Power / select button
7.
Button for opening battery cover
8.
Battery cover
Sensor module
9.
10.
Handle of sensor module
11.
Batteries
12.
USB cable connector
13.
Connection for hand strap
14.
Mouthpiece receptacle
15.
Mouthpiece
16.
Breath funnel (for tests in fast mode)
This way you receive results that are as accurate as possible:
A)
Wait at least 15 minutes after your last sip of alcohol.
Do not eat, drink or smoke during that time.
B)
Do not open the casing.
C)
Do not keep the breathalyzer in cold, hot, moist or
badly ventilated places.
D)
Keep the breathalyzer out of the reach of children.
E)
Conduct three inlet tests after long storage before conducting
the actual alcohol test.
F)
Do not drop the breathalyzer.
G)
Change the battery if it is depleted. Diminished battery capacity
leads to incorrect test results.
H)
Do not conduct tests in loud (sound waves disrupt the sensor),
drafty or poorly ventilated places.
I)
After a test result of 1.000 per mill (per mill = EU setting) / 0.100% BAC
(% BAC = US setting) / 0.500 mg/L (mg/L = UK setting) or
higher the sensor needs rest time. In this case wait at least three
minutes before starting the next test.
J)
Do not conduct tests in very cold, very hot or an unventilated,
smoky ambience.
K)
Store the breathalyzer in a way to avoid penetration
by dust or liquids.
L)
Blowing too hard may cause damage to the sensor as saliva
flows into it.
Preparation for the test:
1. Installing a battery
Push the button on the back of the device to the right side. Pull the battery cover
downwards and remove the battery compartment cover. A new device is delivered
with already included batteries and you first have to remove the small paper inside
the battery compartment (this paper prevents the 2 inserted batteries from dischar-
ging). Close the battery compartment cover again.
2. Inserting a mouthpiece
Insert a mouthpiece into the port on top of the device. Mouthpieces can be used
with orientation to the right or left side. Ensure that the mouthpiece securely fits.
For sanitary and accuracy reasons it is highly recommended to use an extra
mouthpiece for every test.