Getting Started

Functional Check

To make a simple functional check of the P5210 High Voltage
Differential Probe, select a source that supplies AC line voltage and
use the following procedure. This procedure verifies a majority of
the circuitry within the probe. (For a complete performance
verification, refer to the service manual that accompanies this
product.)
1. Use the installation procedure starting on page 9 to connect the
output of the P5210 High Voltage Differential Probe to a
measurement instrument.
2. Connect the inputs, set the range, and perform the check as each
line of Table 1 indicates.
Table 1: Functional check
Range
Mode
setting
Differential
100X or
1,000X
Common
100X or
Mode
1,000X
This completes the functional check procedure.
10
Input 1
Input 2
(+ or –)
(– or +)
Hot
Ground or
Neutral
Hot (same
connection)
Check
Measurement instrument
displays or indicates the line
voltage
No signal
P5210 Instructions