6.2. Measuring DC/AC Current
WARNING!
*
Do not apply voltage directly across the terminals. You must
connect the meter in series with the circuit.
*
The mA terminal is fused. A severe fire hazard and short
circuit danger exists if you apply a voltage with high-current
capability to this terminal. The meter can be destroyed under
such conditions.
To measure current, break the circuit and connect the probes to
two circuit connection points. Never connect the probes across a
voltage source in parallel. Doing so can blow the fuse or damage
the circuit under test.
Note:
The maximum input current is 600 mA. In these ranges
excessive current blows the fuse, which you must replace.
-
-56-
+