a = Up to three letters and/or numbers or blank (not critical to safety)
b = Up to
two
letters and/or numbers or blank (not critical to safety)
c = Up to five letters and/or numbers or blank (not critical to safety)
Specific Conditions of Use
1. The PMA.Evolution shall only be used with the
YCO14-Z Ex-Link
Converter
2. The protective ground connectors on the PMA.Evolution and the YCO14-Z
Ex-Link
Converter shall be connected
2
.
together using a potential equalisation conductor with a cross-sectional area of at least 4 mm
3. The
CAT6
cable used between the PMA.Evolution and the YCO14-Z
Ex-Link
Converter shall be obtained from
Sartorius.
Equipment Ratings:
YCO14-Z
Ex-Link
Converter
Associated intrinsically safe apparatus for connection to Class I, Division 1, Groups C and D hazardous
locations in accordance with drawing 2021459, Associated Intrinsically Safe for connection to Class I,
Zone 1, Group IIB hazardous locations in accordance with drawing 2021459. Ambient temperature range
0°C to 40°C
PMA.Evolution EVO1Xab-c
PMA.Evolution LAB1Xab-c
Intrinsically Safe for Class I, Division 1, Groups
C and
D, hazardous locations in accordance with drawing
2021459, Intrinsically Safe for Class I, Zone 1, Group IIB hazardous locations in accordance with drawing
2021459. Temperature class T4 at an ambient temperature of 0°C to 40°C.
FM Approved for:
Sartorius Lab Instruments GmbH &
Co.
KG
Goettingen, Germany
To verify the availability of the Approved product, please refer to www.approvalguide.com
FM Approvals HLC 5/13